CANLI
Yükleniyor Veriler getiriliyor…
SCI-Expanded JCR Q2 Özgün Makale Scopus
Investigation of the effect of thyroid collar, radiation safety glasses, and lead apron on radiation dose in cone beam CT
Dentomaxillofacial Radiology 2025 Cilt 54 Sayı 3
Scopus Eşleşmesi Bulundu
3
Atıf
54
Cilt
231-238
Sayfa
Scopus Yazarları: Derya Icoz, Osman Vefa Gul
Özet
Objectives: Due to the increasing use of cone-beam CT (CBCT) in dentistry and considering the effects of radiation on radiosensitive organs, the aim of this study was to investigate the effect of shielding on absorbed dose of eyes, thyroid, and breasts in scans conducted with different parameters using 2 different fields of view (FOV). Methods: Dose measurements were calculated on a tissue-equivalent female phantom by repeating each scanning parameter 3 times and placing at least 2 thermoluminescent dosimeters (TLD) on each organ, with the averages then taken. The same CBCT scans were performed in 2 different FOV with shielding including thyroid collar, radiation safety glasses, and lead apron and without shielding. The differences between them were analysed statistically. Descriptive statistics and the Wilcoxon test were used for data analysis. Results: The difference between measurements with and without shielding was statistically significant for all scans (P <. 001). The dose reduction associated with the use of shielding ranged from 26.81% to 52.95%. The dose related to the FOV has shown a significant increase, ranging from 8.30% to 623.54%, due to both the variation in the area affected by the primary beam on the organs and changes in the amount of radiation. Conclusion: There are significant differences in the absorbed dose depending on shielding and FOV usage. Therefore, the CBCT imaging protocol should be optimized by the operator, and special attention should be paid to the use of thyroid collars and radiation safety glasses, considering their effects on image quality.
Anahtar Kelimeler (Scopus)
cone-beam CT radiation dose radiation dosimetry radiation protection

Anahtar Kelimeler

cone-beam CT radiation dose radiation dosimetry radiation protection

Makale Bilgileri

Dergi Dentomaxillofacial Radiology
ISSN 0250-832X
Yıl 2025 / 3. ay
Cilt / Sayı 54 / 3
Sayfalar 231 – 238
Makale Türü Özgün Makale
Hakemlik Hakemli
Endeks SCI-Expanded
JCR Quartile Q2
Yayın Dili İngilizce
Kapsam Uluslararası
Toplam Yazar 2 kişi
Erişim Türü Basılı+Elektronik
Erişim Linki Makaleye Git
Alan Sağlık Bilimleri Temel Alanı Sağlık Fiziği

YÖKSİS Yazar Kaydı

Yazar Adı İÇÖZ DERYA,GÜL OSMAN VEFA
YÖKSİS ID 8544036

Metrikler

Scopus Atıf 3
JCR Quartile Q2
Yazar Sayısı 2