Background: The purpose of this in-vitro study was to evaluate the repeatability and interdevice reliability of two identical spectrophotometers with different software. Methods: Twenty-six shade tabs of a shade guide (Vita 3D-Master, Vita Zahnfabrik) were measured with two identical spectrophotometers (Vita EasyShade V) with different software (v509i: Device A; v505p: Device B), and CIELab values were recorded (n=260). The color difference (∆E<inf>00</inf>) values between measurements of spectrophotometers were calculated. For repeatability, the 3M2 shade tab of the same shade guide was measured 50 times with each spectrophotometer. The matching accuracy rate of the devices with shade guide was calculated. Interdevice reliability and differences between devices were calculated using intraclass correlation coefficient (ICC), Cronbach's alpha, and Paired T-test (α=0.05). Results: Interdevice reliability was found to be high for L*, a*, and b* parameters (ICC>0.99). The ∆E<inf>00</inf> values of the shade tabs range from 0.21 to 1.78, which were lower than the acceptability threshold (∆E<inf>00</inf><1.8). The mean L* value of Device A was significantly higher than that of Device B (P<0.001), while the mean a* and b* values were significantly lower (P<0.001). The repeatability is excellent for a* and b* parameters (SD<0.1), however, slightly lower for the L* (0.15<SD<0.25) in both devices. The matching accuracy rates of both devices were calculated as 34.6 %. Conclusion: The reliability and repeatability of two identical spectrophotometers were found to be high, and the color difference values were found to be at an acceptable level. However, both devices can display different L*, a*, and b* values for the same shade guide.
Anahtar Kelimeler (Scopus)
Color
Equipment failure analysis
Reliability
Repeatability
Spectrophotometry
Anahtar Kelimeler
Color
Equipment failure analysis
Reliability
Repeatability
Spectrophotometry
Makale Bilgileri
Dergi
Selcuk Dental Journal
ISSN
2148-7529
Yıl
2021
/ 12. ay
Cilt / Sayı
8
/ 3
Sayfalar
605 – 610
Makale Türü
Özgün Makale
Hakemlik
Hakemli
Endeks
TR DİZİN
TEŞV Puanı
27,00
Yayın Dili
Türkçe
Kapsam
Ulusal
Toplam Yazar
3 kişi
Erişim Türü
Elektronik
Erişim Linki
Makaleye Git
Alan
Sağlık Bilimleri Temel Alanı
Protetik Diş Tedavisi
YÖKSİS Yazar Kaydı
Yazar Adı
YILMAZ SAVAŞ TUBA, SAVAŞ ABDULHALUK, TUNÇDEMİR ALİ RIZA
YÖKSİS ID
6055675
Hızlı Erişim
Metrikler
TEŞV Puanı
27,00
Yazar Sayısı
3